Damage estimator, a possible way to predict the failure of hip and knee endoprosthesis

Lucian Capitanu (Tribology Department, Institute of Solid Mechanics of the Romanian Academy, Bucharest, Romania)
Virgil Florescu (Institute of Civil Engineering, Bucharest, Romania)
Liliana-Laura Badita (Micro/Nano Mechatronic Technologies Department, National Institute for Research and Development in Mechatronics and Measurement Technique, Bucharest, Romania)

Industrial Lubrication and Tribology

ISSN: 0036-8792

Article publication date: 9 January 2017

137

Abstract

Purpose

The purpose of this study was to realize finite element simulation in order to dynamically determine the area of the contact, the contact pressure and the strain energy density (identified as a damage function) for three different activities – normal walking, ascending stairs and descending stairs – that could be considered to define the level of the activity of the patient.

Design/methodology/approach

The finite element model uses a modern contact mechanism that includes friction between the metallic femoral condyles or femoral head (considered rigid) and the tibial polyethylene insert or acetabular cup (considering a non-linear behaviour).

Findings

For all three activities, the finite element analyses were performed, and a damage score was computed. Finally, a cumulative damage score (that accounts for all three activities) was determined, and the areas where the fatigue wear is likely to occur were identified.

Originality/value

A closer look at the distribution of the damage score reveals that the maximum damage is likely to occur not at the contact surface, but in the subsurface.