To read this content please select one of the options below:

Composite coating structure in an implantable electronic device

Kati Kokko (Department of Electronics, Tampere University of Technology, Tampere, Finland)
Hanna Harjunpää (Department of Biomedical Engineering, Tampere University of Technology, Tampere, Finland)
Pekka Heino (Department of Electronics, Tampere University of Technology, Tampere, Finland)
Minna Kellomäki (Department of Biomedical Engineering, Tampere University of Technology, Tampere, Finland)

Soldering & Surface Mount Technology

ISSN: 0954-0911

Article publication date: 26 June 2009

287

Abstract

Purpose

The purpose of this paper is to investigate the influence of composite coating structure on the reliability of adhesive flip chip joints. The need for conformal coating is considered, especially for medical applications, and medical sterilization is also considered.

Design/methodology/approach

Two test lots were assembled and one of them was sterilized using gamma sterilization. Both test lots were coated first with epoxy and then with Parylene C, resulting in a composite coating structure. The reliability was studied using a constant humidity test and the failure analysis was performed with cross‐sections and scanning electron microscopy analysis. These results were compared to earlier research results on conformal coatings.

Findings

The reliability of both test lots proved to be good. The composite coating structure shields the joints from humidity and improves the reliability compared to non‐coated test samples. When the conformal coating was compared to the pure Parylene C coated test lot, the reliability was almost the same. This leads to the conclusion that the epoxy layer in the composite coating structure has no value when long‐term reliability is considered. Gamma sterilization does not greatly affect reliability. The epoxy coating under the Parylene C layer cracked during reliability testing.

Originality/value

The paper shows the influence of composite coating structure on the reliability of adhesive flip chip joints, particularly important in medical applications.

Keywords

Citation

Kokko, K., Harjunpää, H., Heino, P. and Kellomäki, M. (2009), "Composite coating structure in an implantable electronic device", Soldering & Surface Mount Technology, Vol. 21 No. 3, pp. 24-29. https://doi.org/10.1108/09540910910970385

Publisher

:

Emerald Group Publishing Limited

Copyright © 2009, Emerald Group Publishing Limited

Related articles